Cl II Ortho VITROS 4600 Chemistry System

Company: Ortho-Clinical Diagnostics
Date of Enforcement Report 1/20/2016
Class lI:

PRODUCT

VITROS 4600 Chemistry System, Catalog Number 6802445, Unique Device Identifier Number 10758750012343, Software Version 3.2 and below; IVD. Product Usage: The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)
Recall Number Z-0642-2016

REASON
Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.

RECALLING FIRM/MANUFACTURER
Ortho-Clinical Diagnostics, Rochester, NY on 11/3/2015. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
Domestic: 157, Foreign: 270

DISTRIBUTION
Nationwide and Internationally.

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

Remote Webmeeting Assessments

SoftwareCPR can provide remote offsite assessments to support virtual offices.  Our consultants can utilize webmeeting tools to walk your teams through assessments such as:

  • 62304 compliance
  • Regulatory submission pre-review
  • Software risk analysis
  • Cybersecurity process and validation
  • Overall ISO 14971 risk management
  • Overview of software regulation with John Murray

Email office@softwarecpr.com
for more info!

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.