Cl II CARESTREAM DRX-EVOLUTION X-Ray system

Company:Carestream Health Inc.
Date of Enforcement Report: 9/2/2015
Class lI:

PRODUCT

CARESTREAM DRX-EVOLUTION X-Ray System Product Usage: The device is a permanently installed diagnostic X-ray system for general radiographic x-ray imaging including tomography. The tomography feature is not to be used for imaging pediatric patients. Recall Number Z-2449-2015

REASON
Carestream became aware of an issue with its DRX Evolution System v5.7 DIRECTVIEW Software v5.6 and v5.7. It has been determined that there is a software defect that impacts image alignment when using the automatic stitching option for long length images. The defect could cause stitching inconsistencies which may go undetected on a radiograph.

RECALLING FIRM/MANUFACTURER
Carestream Health Inc, Rochester, NY on 8/7/2015 Voluntary: Firm Initiated recall is ongoing

VOLUME OF PRODUCT IN COMMERCE
Domestic: 76 units, Foreign: 160 units

DISTRIBUTION
Nationwide and Internationally

___________________________________

Upcoming Training

62304, FDA, and Emerging Standards for Medical Device and HealthIT
Instructors:  Brian Pate, John F. Murray, Jr
Location: Sunnyvale, CA, USA
Dates:  February 4-6, 2020
Registration Link

Receive $300 discount with Premium-Individual subscription purchase (or $333 per person for Premium-Company subscription)!  Email training@softwarecpr.com
to receive discount

QSS Software Validation
Planned Instructors:  Brian Pate, John F. Murray, Jr
Location: Boston, MA, USA
Dates:  June 2-4, 2020
For info on this course, email training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.