Cl II SIEMENS Uroskop Omnia Max

Company:Siemens Medical Solutions USA, Inc.
Date of Enforcement Report: 6/3/2015
Class lI:

PRODUCT

SIEMENS Uroskop Omnia Max ; a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: Querying and retrieving patient history information and/or previous diagnosis and images from other modalities, including X-ray examinations of the urogenital area, Ultrasound examinations, Endourological interventions, Percutaneous interventions, Laparoscopy, Application of fistula, Simple procedures, Extracorporeal shock wave lithotripsy, Uroflow/urodynamics, Pediatric radiological and therapeutic applications.
Recall Number Z-1670-2015

REASON
Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.

RECALLING FIRM/MANUFACTURER
Siemens Medical Solutions USA, Inc, Malvern, PA on 4/13/2015 Voluntary: Firm Initiated recall is ongoing

VOLUME OF PRODUCT IN COMMERCE
133

DISTRIBUTION
Nationwide

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.