1 Software Version 2.8 & Below, Cl II VITROS 5

Company:Ortho-Clinical Diagnostics
Date of Enforcement Report: 5/6/2015
Class lI:

PRODUCT

Software Version 2.8 & Below on VITROS 5,1 FS Chemistry Systems, Catalog Number 6801375, Global Trade Item Number 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number 6801890, Global Trade Item Number 10758750001644; IVD. Intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides and VITROS Chemistry Products MicroTip Reagents.
Recall Number Z-1521-2015

REASON
Software Anomaly; It is possible for the device to process samples with a cartridge other than the intended cartridge, potentially leading to erroneous patient results. If this anomaly occurred previously, an indication would be a series of results that were similar across multiple samples (i.e., results could be believable or outside of the Reportable Range for the intended assay).

RECALLING FIRM/MANUFACTURER
Ortho-Clinical Diagnostics, Rochester, NY on 3/10/2015 Voluntary: Firm Initiated recall is ongoing

VOLUME OF PRODUCT IN COMMERCE
VITROS 5.1 System: Domestic – 909; Foreign – 1250; 5,1 Refurbished: Domestic – 66 units, Foreign – 152

DISTRIBUTION
Nationwide and Internationally

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.