Cl II ACCU-CHEK Connect Diabetes Management App

Company:Roche Diabetes Care, Inc.
Date of Enforcement Report: 4/8/2015
Class lI:

PRODUCT

ACCU-CHEK Connect Diabetes Management App; Instruction Manual Designed to transfer data for diabetes management. Recall Number Z-1369-2015

REASON
Roche Diabetes Care has become aware of an issue with the Accu-Chek Connect diabetes management app that could potentially lead to inaccurate bolus advice being provided to the user. A thorough investigation of the situation revealed that this issue may occur if the user changes the screen orientation of the phone from portrait to landscape or vise versa while looking at the Bolus Advisor or Carbohydrate Entry screens.

RECALLING FIRM/MANUFACTURER
Roche Diabetes Care, Inc.Indianapolis, IN on 10/30/2014 Voluntary: Firm Initiated recall is ongoing

VOLUME OF PRODUCT IN COMMERCE
The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.

DISTRIBUTION
There are no US distributors as the affected device is not marketed in the US. The affected device was distributed in Germany, Italy and South Africa. The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.

___________________________________

Software Verification

Do you have gaps with IEC 62304 compliance or with FDA expectations for software verification?  We can help you quickly remediate gaps with unit verification including unit testing, code review, and static analysis, as well as integration and system verification testing.  Call us at 781-721-2921 or email office@softwarecpr.com to start a conversation.  We can provide estimates of cost and support your development tools.

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.