Company:Radiometer America Inc
Date of Enforcement Report: 2/11/2015
 Class lI:
PRODUCT
 
 ABL90 FLEX analyzer, Model number 393-090. A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood.  Recall Number	Z-1046-2015
REASON
 The ABL90 analyzer does not always use the most recent calibration data to calculate patient results. This can in some cases lead to a biased patient result.
RECALLING FIRM/MANUFACTURER
 Radiometer America Inc., Westlake, OH  on 12/82014 Voluntary: Firm Initiated recall is ongoing
VOLUME OF PRODUCT IN COMMERCE
 5002 units
DISTRIBUTION
 Nationwide and Internationally