Company:TeraRecon, Inc.
Date of Enforcement Report: 2/18/2015
Class lI:
PRODUCT
Aquarius iNtuition Client Viewer. Findings Workflow module, RECIST 1.1: Picture Archiving and Communications System; Findings Workflow Modules 4.4.11.82.6784, 4.4.11.116.7134, 4.4.11.144.7589. A fully-configured iNtuition system is capable of various image processing and visualization functions, including basic features and advanced post processing modules. The system can be configured as a server with some, all, or none of its optional features disabled. The intended use of the device is to provide solutions to various medical image-analysis and viewing problems, which come about as modalities generate more and more images. It also supports image distribution over networks, and is DICOM compliant.
Recall Number Z-1070-2015
REASON
Software anomaly related to RECIST1.1 target lesion evaluation criteria in Findings Workflow Module within the Aquarius iNtuition Client Viewer..
RECALLING FIRM/MANUFACTURER
TeraRecon, Inc., Foster City, CA on 1/21/2015 Voluntary: Firm Initiated recall is ongoing
VOLUME OF PRODUCT IN COMMERCE
91
DISTRIBUTION
Nationwide and Internationally