Company:Ortho-Clinical Diagnostics
Date of Enforcement Report: 1/21/2015
Class lI:
PRODUCT
VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), Catalog Number 6802445, IVD — Ortho Clinical Diagnostics. The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods).
Recall Number Z-0968-2015
REASON
Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
RECALLING FIRM/MANUFACTURER
Ortho-Clinical Diagnostics, Rochester, NY on 11/19/2014 Voluntary: Firm Initiated recall is ongoing
VOLUME OF PRODUCT IN COMMERCE
339 Total: USA – 102, Foreign – 237
DISTRIBUTION
Nationwide and Internationally