GE Centricity Enterprise Archive Cl II

Company: GE Healthcare It.
Date of Enforcement Report 12/31/2014
Class lI:

PRODUCT

Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. DICOM devices (e..g. modalities, workstations) Communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction that other systems have with the Centricity Enterprise archive. The intended users for Centricity Enterprise Archive are service personnel and system administrators and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-1 profiles (published by IHE).
Recall Number Z-0833-2015

REASON
GE Healthcare has become aware of a potential data loss issue associated with empty path names resulting from HL-7 updates when the cache disk is almost full..

RECALLING FIRM/MANUFACTURER
GE Healthcare It, Barrington, IL 4/19/2013 Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
115 units

DISTRIBUTION
Nationwide

___________________________________

Upcoming Training

62304, FDA, and Emerging Standards for Medical Device and HealthIT
Instructors:  Brian Pate, John F. Murray, Jr
Location: Sunnyvale, CA, USA
Dates:  February 4-6, 2020
Registration Link

Receive $300 discount with Premium-Individual subscription purchase (or $333 per person for Premium-Company subscription)!  Email training@softwarecpr.com
to receive discount

QSS Software Validation
Planned Instructors:  Brian Pate, John F. Murray, Jr
Location: Boston, MA, USA
Dates:  June 2-4, 2020
For info on this course, email training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.