Customed Inc.

Customed Inc.
Product: medical devices, including convenience packs for surgical procedures
Date: 1/9/2015

Failure to validate the defined user needs and intended uses of the (b)(4) used as part of the quality system according to an established protocol, as required by 21 CFR 820.70(i). Specifically, your firm implemented (b)(4) to track product inventory (incoming materials, finished/released products and quarantine items), however the software was not validated. In addition, the status of lots of finished products and incoming materials in the warehouse failed to match the (b)(4) system inventory.

Your firm?s response dated August 21, 2014, is not adequate. Your firm initiated CAPA (b)(4) to prepare a software validation protocol, however, you failed to initiate a corrective action to identify and address pertinent quality system and production software that have not been appropriately validated. Further, your firm failed to provide the executed (b)(4) validation protocol and summary report for review.

FDA District Office:San Juan Office

Upcoming Training

62304, FDA, and Emerging Standards for Medical Device and HealthIT
Instructors:  Brian Pate, John F. Murray, Jr
Location: Sunnyvale, CA, USA
Dates:  February 4-6, 2020
Registration Link

Receive $300 discount with Premium-Individual subscription purchase (or $333 per person for Premium-Company subscription)!  Email training@softwarecpr.com
to receive discount

QSS Software Validation
Planned Instructors:  Brian Pate, John F. Murray, Jr
Location: Boston, MA, USA
Dates:  June 2-4, 2020
For info on this course, email training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.