BL14334 Cl II Bausch & Lomb Stellaris PC w/ Laser

Company: Bausch & Lomb Incorporated Rochester, NY
Date of Enforcement Report 12/31/2014
Class lI:

PRODUCT

Bausch & Lomb Stellaris PC w/ Laser Upgrade BL14334 (Stellaris PC w v4 Software) with BL2272 (Laser Upgrade kit) Stellaris PC Vision Enhancement System with integrated laser module. System is used for the surgical treatment/removal of cataractous lens, vitreous removal and certain treatments of the posterior portion of the eye.
Recall Number Z-0864-2015

REASON
A software anomaly was identified where the system may automatically transition from “Standby” to “Ready to Fire” state while in surgical use.

RECALLING FIRM/MANUFACTURER
Bausch & Lomb Inc., Saint Louis, MO on 11/4/2014. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
187

DISTRIBUTION
Nationwide and Internationally

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

Upcoming Public Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software training course in Boise, Idaho. 3-Day course.

October 13-15, 2026

Limited Early Bird Registration: $995 / person
Non refundable but transferable.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.