Cl II Varian 4D Integrated Treatment Console

Company: Varian Medical Systems, Inc.
Date of Enforcement Report: 9/24/2014
Class lI:

PRODUCT

4D Integrated Treatment Console (4DITC), versions 10 and 11. 4DITC provided assistance for accurate treatment delivery by monitoring linear accelerator parameters and preventing the radiation therapy device from commencing irradiation when any parameter is out of conformance with the treatment plan. Recall Number Z-2654-2014

REASON
Due to user error, the 4DITC can improperly allow users to clear a safety interlock imposed by the 4DITC on behalf of Varian’s Bar Code Conical Verification system from the BrainLab’s ExacTract console when it is used with BCCV.

RECALLING FIRM/MANUFACTURER
Varian Medical Systems, Inc., Palo Alto, CA on 8/18/2014 Voluntary: Firm Initiated recall is ongoing

VOLUME OF PRODUCT IN COMMERCE
512

DISTRIBUTION
Nationwide and Internationally.

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.