Draeger Infinity Acute Care System Cl II

Company: Draeger Medical Systems, Inc. .
Date of Enforcement Report:1/22/2014
Class lI:

PRODUCT

Draeger Infinity Acute Care System Monitoring Solution Recall Number Z-0714-2014

REASON
After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

RECALLING FIRM/MANUFACTURER
Draeger Medical Systems, Inc. on 12/10/2013. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
439

DISTRIBUTION
Nationwide

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.