Cl II Varian Plan Parameters workspace

Company: Varian Medical Systems, Inc.
Date of Enforcement Report:1/1/2014
Class lI:

PRODUCT

Plan Parameters workspace within Eclipse Treatment Planning System Version 11 and 13; ARIA Radiation Oncology versions 11 and 13 Acuity, versions 11 and 13. Manufactured by Varian Medical Systems Palo Alto, CA. Used in radiation therapy simulation.
Recall Number Z-0552-2014

REASON
An anomaly has been traced to a software deficiency within the Plan Parameters workspace. When a dose relevant parameter is changed and “do no clear MUs and Reference Point dose” is selected, the Monitor units may change.

RECALLING FIRM/MANUFACTURER
Varian Medical Systems, Inc., Palo Alto, CA on 11/18/2013. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
6701

DISTRIBUTION
Nationwide andI Internationally

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

Upcoming Public Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software training course in Boise, Idaho. 3-Day course.

October 13-15, 2026

Limited Early Bird Registration: $995 / person
Non refundable but transferable.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.