Cl ll Canon Full Auto Ref-Keratometer RK-F2

Company: Canon Inc.
Date of Enforcement Report: 6/5/2013
Class ll:

PRODUCT

Canon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA. Product Usage: This device is intended to be used to measure the refractive power and the radius of corneal curvature of the human eye. Recall Number Z-1404-2013

REASON
Canon Inc. had received 17 claims from our foreign sales companies between October to November 2012 that measurement value of RK-F2s refractive power (SPH value) sometimes indicated abnormal value which occurs approximately 5% in probability.

RECALLING FIRM/MANUFACTURER
Canon Inc., Tokyo, Japan on 12/5/2012. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
384 units

DISTRIBUTION
Nationwide and Internationally

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.