FDA Mobile App Compliance Letter

Puget FDA Mobile App Compliance Letter
Product:Letter to Biosense Technologies Private Limited concerning the uChek Urine Analyzer
Date: 5/21/2013

It has come to our attention that you are currently marketing the uChek Urine analyzer, which is intended for use with Siemens Multistix SG10, Siemens Multistix SG, Siemens Uristix, Bayer Diastix, and Bayer Keto-Diastix reagent strips for the qualitative and semi-quantitative determination of urine analytes including glucose, urobilinogen, pH, ketone, blood, protein, bilirubin, nitrite, leukocyte, and specific gravity. The uChek Urine analyzer appears to meet the definition of a device as that term is defined in section 201(h) of the Federal Food Drug and Cosmetic Act.
Please note that though the types of urinalysis dipsticks you reference for use with your application are cleared, they are only cleared when interpreted by direct visual reading. Since your app allows a mobile phone to analyze the dipsticks, the phone and device as a whole functions as an automated strip reader. When these dipsticks are read by an automated strip reader, the dipsticks require new clearance as part of the test system. Therefore, any company intending to promote their device for use in analyzing, reading, and/or interpreting these dipsticks need to obtain clearance for the entire urinalysis test system (i.e., the strip reader and the test strips, as used together). For an example of this type of device system, and a summary of the type of data used to support clearance of the system, see the decision summary for k111221 (http://www.accessdata.fda.gov/cdrh_docs/reviews/K111221.pdf1).

We have conducted a review of our files, and have been unable to identify any Food and Drug Administration (FDA) clearance number for the uChek Urine analyzer. We request that you provide us with the FDA clearance number for the uChek Urine analyzer. If you do not believe that you are required to obtain FDA clearance for the uChek Urine analyzer, please provide us with the basis for that determination. Please provide the requested information within thirty (30) business days.

We have assigned a unique document number that is cited above. The requested information should reference this document number and should be submitted to:

James L. Woods, WO66-5688
Deputy Director
Patient Safety and Product Quality
Office of In Vitro Diagnostics and Radiological Health
10903 New Hampshire Avenue
Silver Spring, MD 20993

If you have questions relating to this matter, please feel free to call Amy Ghering at 301-796-1656 or at Amy.Ghering@fda.hhs.gov, or log onto our web site at www.fda.gov2 for general information relating to FDA device requirements.

Sincerely yours,

James L. Woods
Deputy Director,
Patient Safety And Product Quality
Office of In Vitro Diagnostics and Radiological Health
Center for Devices and Radiological Health

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Being Agile & Yet Compliant (Public or Private)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for scheduling!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

2-days onsite (4 days virtual) with group exercises, quizzes, examples, Q&A.

Instructors: Mike Russell, Ron Baerg

Next public offering: March 7 & 28, 2024

Virtual via Zoom

Registration Link:

Register Now

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.