Cl ll Toshiba Infinix-i X-Ray System.

Company:Toshiba American Medical Systems Inc .
Date of Enforcement Report4/3/2013
Class ll:

PRODUCT

Toshiba Infinix-i X-Ray System. A radiographic and fluoroscopic x-ray system for use in diagnostic and angiographic procedures. Recall Number Z-0994-2013

REASON
Lost and incorrect images. When using the “F-Rec” fluoroscopy image acquisition recording option in the “manual mode” an infrequent software error may cause images from another study to be displayed on the monitor. The acquired images may be lost requiring additional image acquisitions to be performed.

RECALLING FIRM/MANUFACTURER
Toshiba American Medical Systems Inc., Tustin, CA, on 11/8/2012. Voluntary: Firm Initiated recall has been terminated.

VOLUME OF PRODUCT IN COMMERCE
95 units

DISTRIBUTION
Nationwide

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.