Company:Intel-GE Care Innovations LLC. 
Date of Enforcement Report 3/6/2013
 Class ll:
PRODUCT
 
 QuietCare-Networked product, Facility Server/model: QC101601; The QuietCare system is intended to detect and analyze residents’ movements.
 Recall Number  Z-0877-2013
REASON
 Care Innovations detected a software anomaly: During the engineering investigation into the cause of why a Wander Alert was not generated, a defect in the implementation of the Wander and Night Motion ADLs was discovered.
RECALLING FIRM/MANUFACTURER
 Intel-GE Care Innovations LLC, Rosevill, CA on 12/15/2012. Voluntary: Firm Initiated recall has been terminated
VOLUME OF PRODUCT IN COMMERCE
 19 units
DISTRIBUTION
 Nationwide
___________________________________