Cl ll DataCare GM

Company: Roche Molecular Systems, Inc.
Date of Enforcement Report 12/19/12
Class ll:

PRODUCT

DataCare GM DataCara GM is an information management system that provides data management functionality for managing blood glucose testing data.
Recall Number Z-0492-2013

REASON
Symantec, a third party software vendor, informed Roche about a defect in their pcAnywhere software (versions 12.6.6 and older), which is a software product for accessing PCs remotely. The defect is a “pre authentication vulnerability” which allows hackers to gain full access to the target system without having to log on. The following Roche products are potentially affected by this issue: AMPLILINK Data Station, MagNA Pure LC, MagNA Pure 96 Instrument IVD, LightCycler 1.2 Instrument, LightCycler 2.0 Instrument, cobas 8000, Instrument Manager, DataCare GM, DataCare POC and COBAS INTEGRA 400/400+ .

RECALLING FIRM/MANUFACTURER
Roche Molecular Systems, Inc. Branchburg, NJ on 6/25/2012. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
479 units

DISTRIBUTION
Nationwide

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.