Company: GE Healthcare, LLC
Date of Enforcement Report 10/31/12
Class ll:
PRODUCT
GE Seno Advantage 1.x, Seno Advantage 2.0, Seno Advantage 2.1 Seno Advantage 1.x, Seno Advantage 2.0, Seno Advantage 2.1: Seno Advantage is a medical image review station that allows easy selection, processing, filming and media interchange of multi-modality images from a variety of diagnosis imaging systems. When interpreted by a trained physician, mammographic images displayed on the high-resolution monitors may be used as an element for diagnosis. Furthermore filmed images from all modalities may also be used as an element for diagnosis. Recall Number Z-0095-2013
REASON
GE Healthcare has become aware of an issue associated with the magnification images acquired on non-GE digital mammography systems and your Seno Advantage Workstation that may impact patient safety. The measurement values provided by the Seno Advantage Workstation may be incorrect when applied to magnification images acquired on non-GE digital mammography systems. If not noticed by the caregiver, this could lead to an overestimate of the size of the breast lesion. The measurement values for GE images are not affected.
RECALLING FIRM/MANUFACTURER
GE Healthcare, LLC, Waukesha , WI on 5/24/2012. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
188
DISTRIBUTION
Nationwide and Internationally
___________________________________