, Cl ll GE Healthcare Advantage Workstations

Company:GE Healthcare, LLC
Date of Enforcement Report 8/22/12
Class ll:

PRODUCT

GE Healthcare Advantage Workstations, Advantage Worksation version 4.2 and 4.3 with Volume Viewer 2 (version 6.4-6.8) provided with Advance Vessel Analysis (AVA) and Advantage Workstation 4.4 with Volume Viewer 2 (version 6.8 or before) or Volume Viewer 3 (version 7.2 -7.3) provided with Advanced Vessel Analysis (AVA), Model 2378698-2 Product Usage: Advantage Workstation version 4.2: The Advantage Workstation 4.2 is a review station, which allows easy selection, review, processing, filming and medial interchange of multi-modality images from a variety of diagnostic imaging systems. When interpreted by a trained physician, filmed images may be used as an element for diagnosis. Advantage Workstation version 4.3/4.4 : The Advantage Workstation 4.3 is a review station, which allows easy selection, review, processing, filming and medial interchange of multi-modality DIACOM images from a variety of diagnostic imaging systems. When interpreted by a trained physician, filmed images of the AW monitor may be used as a basis for diagnosis, except in the case of mammography. Z-2203-2012

REASON
GE Healthcare is informing you of incorrect measurements when using the Aorta analysis protocol or customized protocols. Until a new software version is installed they are asking you not to use the aorta analysis protocol or the customized protocols for an aneurysm. A second issue was also discovered internally. When loading PET data into the Volume Viewer, if slices are missing in a PET series (i.e., from a networking error, or operator error), the Volume Viewer application will interpolate from neighboring slices to fill these gaps when loading the series without any warning. Thus, if a small lesion is located in any of the missing slices, it may not appear on the interpolated slices.

RECALLING FIRM/MANUFACTURER
GE Healthcare, LLC., Waukesha WI on 5/24/2012. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
1055

DISTRIBUTION
Worldwide

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.