GE Advantage CT/PET Fusion Software Cl ll

Company:GE Healthcare, LLC
Date of Enforcement Report 7/18/12
Class ll:

PRODUCT
GE Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator 2 software. (Version 4.0, 4.1,or 4.2) The Advantage CT/PET Fusion clinical application provides easy comparison of three dimensional (3D) anatomical images from Computed Tomography (CT) and functional images from Positron Emission Tomography (PET). Z-1960-2012

REASON
GE Healthcare has become aware of a potential safety issue due to incorrect alignment of fused CT and PET images associated with the CT/PET fusion capabilities of your Volume Viewer and CT/PET Fusion application. Volume Viewer allows two studies (one PET and one CT) that are registered in the same spatial domain to be viewed simultaneously in fused mode. The two images should be aligned to the same precision as the scanner. Assuming the two images are acquired with a calibrated machine and there is no patient movement between the CT and PET acquisitions, the two images should be visually aligned to a precision of less than one half PET voxel. However, the problem associated with this version of software may introduce a shift of up to 1.5 PET voxels, which can be as much as 8mm with certain zoom factors.

RECALLING FIRM/MANUFACTURER
GE Healthcare, LLC, Waukesha, WI on 5/24/2012. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
70 units

DISTRIBUTION
Nationwide and Internationally

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.