Softlab Lab Info System, Cl ll SCC Soft Computer

Company:Steris Corp
Date of Enforcement Report 12/15/11
Class ll:

PRODUCT
SCC Soft Computer, Softlab Laboratory information System with DORA versions: 2.0.0.4 with Hot Fix 1.2808, 1.2809, 1.2956, or 1.3476; 2.0.0.5; 2.0.0.6; 2.0.0.7 with Hot Fix 1.6595; and 2.1.0.3. Recall #Z-0278-2012.

REASON
Potential for patient data to be sent to the wrong patient record. A software change affected how the Softlab handles recovery from certain types of table access conflicts. When such conflicts occur, this function has the potential to cause faulty writing to the database.

RECALLING FIRM/MANUFACTURER
Recalling Firm: Steris Corp., Mentor, OH, through SCC Soft Computer’s proprietary Task Management System on March 22, 2011.
Manufacturer: Steris Corp., Montgomery, AL. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
12 units

DISTRIBUTION
Nationwide and Canada

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.