Cl ll Plum A+ and A+3 Infustion Pumps

Company: Hospira Inc.
Date of Enforcement Report 10/26/11
Class ll:

PRODUCT
1) Plum A+ Single Channel Infusion Pumps; The pumps were sold under the following configurations: a) list 11971 b) list 11973 – software version 10.3 c) list 12391 – software version 11.3 d) list 20679 – Hospira MedNet Software e) list 20792 – Driver The Plum A+ dual-line volumetric infusion system designed to meet the growing demand for hospital-wide, as well as alternate site and home healthcare, standardization. With their primary line, secondary line, and piggyback fluid delivery capability, the Plum A+ is suited for a wide range of medical/surgical and critical care applications. Full compatibility with LifeCare Plum Series administration sets and accessories and the LifeShield needleless protection systems makes the Plum A+ a convenient and cost-effective infusion system. Recall # Z-3284-2011

2) Plum A+3 Triple Channel Infusion Pumps; the pumps were sold under the following configurations: a) list 12348 – software version 10.3; b) list 12618 – software version 11.3; c) list 20678 – Hospira MedNet software The Plum A+3 is a dual-line triple channel volumetric infusion system designed to meet the growing demand for hospital-wide, as well as alternate site and home healthcare, standardization. With their primary line, secondary line, and piggyback fluid delivery capability, the Plum A+3 is suited for a wide range of medical/surgical and critical care applications. Full compatibility with LifeCare Plum Series administration sets and accessories and the LifeShield needleless protection systems makes the Plum A+3 a convenient and cost-effective infusion system. Recall # Z-3285-2011;

3) Plum A+ Hyperbaric Single Channel Infusion Pumps; list 11005. Recall # Z-3286-2011

REASON
Hospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the clinician has not engaged the clamp prior to opening the cassette door, unrestricted flow may occur..

RECALLING FIRM/MANUFACTURER
Recalling Firm: Hospira Inc., Lake Forest, IL, by letter dated September 16, 2011.
Manufacturers: Hospira Holdings de, Costa Rica Ltd., La Aurora de Heredia,
Costa Rica;
Manufacturer: Hospira Inc., Morgan Hill, CA. Firm initiated recall is ongoing.|

VOLUME OF PRODUCT IN COMMERCE
290,217 pumps

DISTRIBUTION
Nationwide and Internationally

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.