Cl ll ABL90 FLEX Portable Blood Gas Analyzer Syst

Company: Radiometer America Inc.
Date of Enforcement Report 6/22/2011
Class ll:

PRODUCT
ABL90 FLEX Portable Blood Gas Analyzer System, Part Number: 393-090, 510 (k) #K092686. All units equipped with software versions: 2.4.1680.25 and all previous versions are subject to recall. The ABL90 FLEX is a portable, automated analyzer that measures pH, blood Gases, electrolytes, glucose, lactate and oximetry in whole blood. The ABL90 FLEX is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. Recall # Z-2112-2011

REASON
The firm determined that the current operating software for their ABL 90 Analyzer does not prevent the further use of the sensor cassette when there is the possibility of iron rich plaque build-up in the reference electrode. If the device operator/user ignores the error message and continues to obtain biased measuring results using the cassette, this could lead to a clinically serious misinterpretation of the results and possible misdiagnosis and/or mal-therapy in the critical patient care setting.

RECALLING FIRM/MANUFACTURER
Recalling Firm: Radiometer America Inc, by letters on February 4, and on February 24th, 2011.
Manufacturer: Radiometer Medical ApS, Bronshoj, Denmark. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
533 units

DISTRIBUTION
Nationwide and Canada

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Contact Me

Fields marked with an * are required

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.