Cl ll Extended Brilliance Workstation

Company: Philips Medical Systems, Inc.
Date of Enforcement Report 4/13/2011
Class ll:

PRODUCT
AExtended Brilliance Workstation (EBW) a component of the Computed Tomography X-Ray System Model Numbers EBW (CT) 728260 & EBW (Portal) 728269. The Extended Brilliance Workplace?? (EBW) is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories. The EBW is also provided as a component of PET/CT and SPECT/CT systems, performing the same function as indicated above (also referred to as the NM EBW). Model Numbers EBW (CT) 728260 & EBW (Portal) 728269. EBW and NM EBW. Recall # Z-1819-2011

REASON
The firm determined that the functional analysis feature in the Comprehensive Cardiac Analysis (CCA) was displaying incorrect values on the polar fields. The calculated results for the myocardial wall thickening was incorrect in that the numbers displayed for “Wall Thickening” actually represent a reduction in wall thickness from ES to ED.

RECALLING FIRM/MANUFACTURER
Philips Medical Systems (Cleveland) Inc., Cleveland, OH, by letter dated September 29, 2010. FDA initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
2,068 units

DISTRIBUTION
Nationwide and Internationally

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Contact Me

Fields marked with an * are required

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.