4000 and 5000 Monitors, Cl ll Dash 3000

Company: GE Healthcare, LLC.
Date of Enforcement Report 2/2/2011
Class ll:

PRODUCT
GE Healthcare Transport Pro Monitor with the CARESCAPE” Patient Data Module. Recall # Z-0363-2011.

REASON
Transport Pro Monitor stops communication with the CARESCAPE Patient Data Module (PDM) after 414 days of continuous run time. This time will equate to different amounts of real time depending on how much the units is actually in service per day. Transport Pro contains an internal timer that is used to control the software and remind users to perform preventive maintenance. When this internal timer rolls-over, the Transport Pro will display an EC1 message. If the Transport Pro is continued to be used without clearing the EC1 message, after additional 19 days of service time, loss of communication between the Transport Pro and a connected CARESCAPE Patient Data Module (PDM) will occur. This results in the loss of the displayed parameter information (e.g. waveforms, numerical data, and alarms). This condition affects all released versions of software.

RECALLING FIRM/MANUFACTURER
GE Healthcare, LLC, Waukesha, WI, by letter dated September 17, 2010.
Manufacturer: General Electric Medical Systems Information Technology, Milwaukee, WI. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
3526

DISTRIBUTION
Nationwide and Internationally

CSV Training Course

Learn FDA expectations for software validation for computer systems, quality system software, manufacturing and production process software, and engineering tools. Email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.