Cl l Baxter HomeChoice and HomeChoice PRO

Company:Baxter Healthcare Corporation.
Date of Enforcement Report 3/4/2010
Class:l

PRODUCT
Baxter Healthcare Corporation, HomeChoice and HomeChoice PRO Automated Peritoneal Dialysis Systems. These systems have been distributed since 1990

REASON
Baxter is conducting a recall of the HomeChoice and HomeChoice PRO because of reports of serious injuries and at least one death associated with increased Intraperitoneal Volume (IIPV), also known as overfill of the abdominal cavity. IIPV can cause serious breathing and heart problems that can result in serious injury or death.

RECALLING FIRM/MANUFACTURER
Baxter Healthcare Corporation, Deerfield, IL

Baxter Public Notice
Baxter is continuing to investigate the causes, fixes, and mitigations associated with IIPV and use of the HomeChoice device. Current mitigations under development and implementation include changes to device labeling and software, which are intended to address issues such as prescription and patient errors, including bypassing alarms. Patients will receive new Patient At-Home Guides, which contain expanded information about IIPV, the symptoms, warnings and cautions, and how to address IIPV, should it occur. Baxter has also developed and is validating a software modification to address this issue and plans to submit a 510(k) to the FDA midyear. The software update will include additional user interface messages and alarms, and change default settings and allowable ranges to reduce risk of excessive accumulation of fluid in the peritoneal cavity.

FDAComments
FDA Comments:

Although Baxter is not removing the HomeChoice and HomeChoice PRO from the market, clinicians should weigh the risks and benefits to continued use of these devices by their patients versus other forms of dialysis therapy. Clinicians should also review the prescription settings for patients who continue to use these devices.

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.