Company:Picis, Inc.
Date of Enforcement Report 12/9/2009
 Class:ll
PRODUCT
 PICIS, CareSuite Family- OR Manager (4235), Preop Manager (4250), Anesthesia Manager (4220), PACU Manager (4245) & Critical Care Manager (1110), patient information systems. Recall # Z-0365-2010
REASON
 Demographic data, most notably allergy and precaution data, can be overwritten with incomplete data or blanks by the interface between non-Picis clinical systems (your “HIS” or “CIS”) and certain Picis applications, if that interface is not configured properly.
RECALLING FIRM/MANUFACTURER
 Picis Inc., Wakefield, MA, by letter on April 8th 2009. Firm initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
 325 units
DISTRIBUTION
 Nationwide, Australia, Austria, Belgium, Denmark, Finland, France, Germany, Italy, Malaysia, Netherlands, Norway, Portugal, Singapore, Spain, Sweden, and United Kingdom