Aria Radiation Oncology, Cl Il Varis

Company:Varian Medical Systems Oncology Systems
Date of Enforcement Report 10/21/2009
Class:ll

PRODUCT
Varis, Aria Radiation Oncology – Version 8.1.15, 8.5.11, and 8.6.07, Model number HIT-INFSYS, The Varis RV function is designed to assist the operator of a radiation therapy device in providing accurate treatment setups for each patient by monitoring set up parameters and preventing the radiation therapy device from commencing irradiation while any parameter is out of conformance with the treatment plan. Additional applications provide various data management and library functions. Recall # Z-1809-2009p>

REASON
Potential mistreatment– due to software issues, retired plans may be used and treated repeatedly.

RECALLING FIRM/MANUFACTURER
Varian Medical Systems Oncology Systems, Palo Alto, CA, by letter dated May 8, 2009. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
1,587 units

DISTRIBUTION
Nationwide and Internationally

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.