Company:American Optisurgical Inc.Date of Enforcement Report 9/16/2009 Class:ll PRODUCT American Optisurgical, Horizon Phacoemulsification System, Model: HZN-2030, for ophthalmic surgery. Recall # Z-1808-2009 REASON Device operation interrupted: A software bug causes Phaco to stop when vacuum goes above 300 mmHg while in Multi Burst mode. Use of this product at the parameters specified above could lead...