Cl Il Brilliance iCT software v2.5.0

Company: Philips Medical Systems
Date of Enforcement Report 7/8/2009
Class:ll

PRODUCT
Brilliance iCT software v2.5.0, Model 728306. The Brilliance iCT is packaged INA carton and the software is contained within the Brilliance CT scanner. The ‘Brilliance iCT’ is a Computed Tomography X-ray System intended to produce cross-section images of the body by computer reconstruction of x-ray transmission data taken at different Ganges and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories. Recall # Z-1582-2009

REASON
Software anomaly. An artifact may appear in brain scans and brain perfusion studies, when using 80kV and a UA, UB, or UC filter with this CT scanning device.

RECALLING FIRM/MANUFACTURER
Philips Medical Systems (Cleveland) Inc, Cleveland, OH, by letters dated December 22, 2008 and January 29, 2009. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
44 units

DISTRIBUTION
DC, CA, GA, IL, IN, KY, NY, MD, MN, OH, OR, MD, Germany, Greece, Israel, Italy, Japan, Netherlands, Norway, Russia, South Korea, Switzerland, Taiwan, Canada, and Thailand

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.