Steris Isomedix Inc

Recipient: Steris Isomedix Inc
Product:sterilizes medical devices.
Date: 9/25/2008

We have reviewed your response dated July 16, 2008, and have concluded that it is inadequate because while you have opened a Correction and Preventative Action (CAPA) to address this issue, your corrective actions did not address any product that was processed prior to the initiation of this CAPA.

2. Failure to establish and maintain adequate procedures to identify all the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems as required by 21 CFR 820.100(a)(3). For example, on 2/15/07, load R070215-07 was processed using revision 10 of the cycle; however, revision 9 was the currently approved version. Your firm was notified by the customer that the load was processed incorrectly. You subsequently opened a CAPA to address this issue and later identified the root cause of the problem as the [redacted] server migration of unapproved cycles into the [redacted] software. However, at the time, you did not initiate any preventative actions to ensure that this problem does not recur.

5. Failure to adequately validate computer software used as part of production, as required by 21 CFR 820.70(i). For example, your [redacted] software, used for the sterilization chambers, was not adequately validated in that unapproved ETO sterilization cycle revisions are migrated from the [redacted] server into the software and are not identified as unapproved cycles. The ETO chamber operators can access these unapproved cycles and run the non validated sterilization cycles without the customers consent.

FDA New England District Office

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.