Company: GE Healthcare Integrated IT Solutions
Date of Enforcement Report 10/8/2008
Class:ll
PRODUCT
GE Centricity AW Suite software version 2.0 and 2.0.1 with Card IQ Xpress Pro or Plus Module; for diagnostic image analysis, Recall # Z-2183-2008
REASON
When using any vessel analysis protocol of the AW Suite Card IQ application, the software can reload saved tracking objects (Save State) incorrectly and display an incorrect vessel label over the restored images. This failure only occurs when users create a custom label in the vessel analysis protocol and include leading or trailing blank characters (space bar) in the custom label.
RECALLING FIRM/MANUFACTURER
GE Healthcare Integrated IT Solutions, Barrington, IL, by letters dated June 11, 2008. Firm initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
207 units
DISTRIBUTION
Nationwide and Internationally