Beckman Synchron and Unicel Class ll

Company: Beckman Coulter Inc
Date of Enforcement Report 5/28/2008
Class:ll

PRODUCT
a) Synchron LX20 Clinical Chemistry System, Part Numbers: 466200,473032,
Recall # Z-1094-2008;
b) Synchron LX20 PRO, Part Number 476100, Recall # Z-1095-2008;
c) Synchron LXi 725, Part Number A08357, Recall # Z-1096-2008;
d) UniCel DxC 600 Part Number: A10405, Recall # Z-1097-2008;
e) UniCel DxC 600i, Part Numbers: A25638, A25639, Recall # Z-1098-2008;
f) UniCel DxC 600 PRO Part Number: A10400, Recall # Z-1099-2008;
g) UniCel DxC 800 Part Number: A10411, Recall # Z-1100-2008;
h) UniCel DxC 800 PRO Part Number: A10407, Recall # Z-1101-2008

REASON
Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors. -If a stall occurs during a GLUm test near the time of sample inject, results may be affected. -Affected GLUm results may be high or low.

RECALLING FIRM/MANUFACTURER
Beckman Coulter Inc, Brea, CA, decided that Customer Notification is not needed. Replacements began April 5, 2007. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
2,599 units

DISTRIBUTION
Nationwide and Internationally

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.