Siemens-charged-particle radiation systems Cl II

Company: Siemens Medical Solutions, USA, Inc
Date of Enforcement Report 5/7/2008
Class:ll

PRODUCT
a) Siemens Coherence AG Therapist, Part No: 5863506, medical charged-particle
radiation therapy system with software version 2.1, Recall # Z-0971-2008;
b) Siemens Coherence Therapist system, Part No: 7339125, medical charged-particle
radiation therapy system with software version 2.1, Recall # Z-0972-2008;
c) Siemens Coherence Impression Therapist system, Part No: 7341410, medical
charged-particle radiation therapy system with software version 2.1,
Recall # Z-0973-2008;
d) Siemens Primeview 3i medical charged-particle radiation therapy system,
Part No: 7341428, with software version 3i, Recall # Z-0975-2008;
e) Siemens AG Therapist 3rd Party V&R medical charged-particle radiation therapy
system, Part No: 7345411, Recall # Z-0975-2008;
f) Siemens Impression Therapist 3rd Party V&R medical charged-particle radiation
therapy system, Part No: 7345429, Recall # Z-0976-2008;
g) Siemens Syngo based WS 3rd Party V&R medical charged-particle radiation
therapy system, Part No: 7345437, Recall # Z-0977-2008;
h) Siemens Coherence Therapist 2.0 medical charged-particle radiation therapy
system, Part No: 8139839, with software version 2.1, Recall # Z-0978-2008;
i) Siemens Primeview 3i 2.0 medical charged-particle radiation therapy system,
Part No: 8139487, with software version 3i 2.0, Recall # Z-0979-2008;
j) Siemens Coherence Therapist 2.1 medical charged-particle radiation therapy system,
Part No: 8147667, with software version 2, Recall # Z-0980-2008;
k) Siemens Primeview 3i 2.1 medical charged-particle radiation therapy system, Part
No: 8147675, with software version 3i, Recall # Z-0981-2008

REASON
Image Orientation Incorrect (Image Shifts) – 1) if the reference image has been calibrated for centering and the reference image is used for patient positioning, the image will shift when the positioning tools are used. 2) This problem occurs only when non-square reference images are used, such as when images are cropped. At this point, the images will refresh, and the center of the reference image will shift.

RECALLING FIRM/MANUFACTURER
Siemens Medical Solutions, USA, Inc., Concord, CA, by letter on October 3, 2007. Firm initiated recall is ongoing

VOLUME OF PRODUCT IN COMMERCE
118 units

DISTRIBUTION
Nationwide and Internationally

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.