BD Biosciences Cytometry Class II

Company: BD Biosciences,
Date of Enforcement Report 3/12/2008
Class:ll

PRODUCT
BD Biosciences MultiSET Flow Cytometry Software Versions 2.1, 2.1.1, and 2.2 with OS X operating systems, Recall # Z-0570-2008

REASON
Inaccurate results: Software error results in inaccurate display result statistics. If the user adjusts the lymph gate or attractors in the Lab Report view without using the Manual Gate function (as described in the system manual), the statistic results will not be updated to match the adjusted gate.

RECALLING FIRM/MANUFACTURER
BD Biosciences, San Jose, CA, by letter on September 14, 2007. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
1,136 units

DISTRIBUTION
Nationwide and Internationally

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.