GE Healthcare Voluson E8 ultrasound scanner Cl II

Company:GE Healthcare
Date of Enforcement Report:11/14/2007
Class:ll

PRODUCT
GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with these software versions; Recall # Z-0113-2008

REASON
When a spectrum of a Pulsed Wave Doppler shows an angle correction unequal to 0 is reloaded from the archive or the image is sent to a DICOM station, this angle correction is mistakenly disregarded resulting in an underestimation of the flow velocities.

RECALLING FIRM/MANUFACTURER
IRecalling Firm: GE Healthcare, Wauwatosa, WI, by letters on July 15, 2007.
Manufacturer: GE Medical Systems, Kretztechnik GmbH & Co. OHG, Zipf, Austria. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
521 units

DISTRIBUTION
Nationwide

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

Upcoming Public Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software training course in Boise, Idaho. 3-Day course.

October 13-15, 2026

Limited Early Bird Registration: $995 / person
Non refundable but transferable.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.