Company:General Electric Medical Systems Information Technology 
Date of Enforcement Report:6/20/2007
 Class:ll
PRODUCT
 GE Healthcare CIC Pro Clinical Information Center, Central Station, Software Version 4.1. (System hardware consists of a processing unit and a monitor), Patient Monitor, Recall # Z-0891-2007
 REASON
 Alarm and Reboot problems: Two separate issues could occur with the CIC Pro system when used with Telemetry. Each issue occurs in software version v4.1 only; A) the system may experience a loss of audible alarms when operating system memory resources reach 170MB and/or B) the system may intentionally or unexpectedly reboot.
 CODE
 All serial numbers of Software Version 4.1
 RECALLING FIRM/MANUFACTURER
 General Electric Medical Systems Information Technology, Milwaukee, WI, by letter on March 13, 2007. Firm initiated recall is ongoing.
 VOLUME OF PRODUCT IN COMMERCE
 5,190 units
 DISTRIBUTION
 Nationwide
