Class II Toshiba American Med Sys Inc.

Company:Toshiba American Med Sys Inc.,
Date of Enforcement Report:5/9/2007
Class:ll

PRODUCT
Nemio XG, ultrasound system, SSA-590A with version 1.5 software, Recall # Z-0771-2007

REASON
Transducer Surface Temperature : It was found during testing that when using the system in the 2DCDI + PW mode, with the Scale Graph function active, that the transducer surface temperature could exceed the 27 degrees Celsius limit as specified in the IEC safety standard..

CODE
Serial #: A1A0642100; A1A0642102; A1B0662183; A1B0662184; A1B0662256; A1B0662257; A1B0662258; A1B0662327; A1B0662328; A1B0662330; A1B0662331; A3A0652115; A3B0662284; A3B0662285; A3B0662286; A3B0662287; A3B0662288; A3B0662289; A3B0662290; A3B0662291; A3B0662292; A3B0662293; A3B0662371; A3B0662372; A3B0662373; A3B0662374; A3B0662375; A3B0662376; A3B0662377; A3B0662378; A3B0662379; A3B0662380; A3B0672415; A3B0672416; A3B0672417; A3B0672418; A3B0672419

RECALLING FIRM/MANUFACTURER
Recalling Firm: Toshiba American Med Sys Inc., Tustin, CA, by letters on December 19, 2006.
Manufacturer: Toshiba Medical Systems Corp., Tokyo, Japan. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
37 units

DISTRIBUTION
Nationwide

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software

ISO 14971 Risk Management (typically with heavy emphasis on software aspect)

IEC 81001-5-1 Cybersecurity Lifecycle

ISO 13485 QMS Requirements & optionally ISO 19011 for leading audits

Being Agile Yet Compliant (AAMI TIR 45)

IEC 29119-1 Software and System Testing

US Classification for Software Devices

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.