Inc., Class II bioMerieux

Company: bioMerieux, Inc,
Date of Enforcement Report:7/8/2006
Class:lI

PRODUCT
OBSERVA R02 (version R02.00.17) computers using PSC 6000 or HHP Barcode Scanners which are connected to BacT/ALERT 3D. Recall # Z-1161-06

REASON
Software problem with scanner inter-character delay which may result in false positive or false negative results after incorrect scanning of bottle IDs.

CODE
Software version R.02.00.17

RECALLING FIRM/MANUFACTURER
Recalling Firm: bioMerieux, Inc., Hazelwood, MO, by telephone and letters on May 22, 2006.
Manufacturer: bioMerieux, Inc., Durham, NC, firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
51 systems

DISTRIBUTION
Nationwide and Internationally

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.