Company: bioMerieux, Inc, 
Date of Enforcement Report:7/8/2006
 Class:lI
PRODUCT
 OBSERVA R02 (version R02.00.17) computers using PSC 6000 or HHP Barcode Scanners which are connected to BacT/ALERT 3D. Recall # Z-1161-06
 REASON
 Software problem with scanner inter-character delay which may result in false positive or false negative results after incorrect scanning of bottle IDs.
 CODE
 Software version R.02.00.17
 RECALLING FIRM/MANUFACTURER
 Recalling Firm:  bioMerieux, Inc., Hazelwood, MO, by telephone and letters on May 22, 2006.
 Manufacturer:  bioMerieux, Inc., Durham, NC, firm initiated recall is ongoing.
 VOLUME OF PRODUCT IN COMMERCE
 51 systems
 DISTRIBUTION
 Nationwide and Internationally