Class II Roche Diagnostics Corp.

Company: Roche Diagnostics Corp.
Date of Enforcement Report:2/1/06
Class:II

PRODUCT
a) Roche brand Accu-Chek Aviva Care Kit;
Catalog number 04528280001. (Distributed
within the U.S.), Recall # Z-0417-06;
b) Roche brand Accu-Chek Aviva Meter Only Kit;
Catalog number 04532279001. (Distributed
within the U.S.), Recall # Z-0418-06;
c) Roche brand Accu-Chek Aviva Bonus Pack;
Catalog number 04528247001. (Distributed
within the U.S.), Recall # Z-0419-06;
d) Roche brand Accu-Chek Aviva Retail
Distributor Pack; Catalog number 04578198001.
(Distributed within the U.S.),
Recall # Z-0420-06;
e) Roche brand Accu-Chek Aviva Bonus Pack
Display; Catalog number 045659066001.
(Distributed within the U.S.)
Recall # Z-0421-06;
f) Roche brand Accu-Chek Aviva Pediatrics Kit;
Catalog number 04593839001. (Distributed
within the U.S.) Recall # Z-0422-06;
g) Roche brand Accu-Chek Aviva Meter (mg/dL);
Catalog number 03532275001. (Distributed
within the U.S.) Recall # Z-0423-06;
h) Roche brand Accu-Chek Aviva Care Kit;
Catalog number 04422058170; – Canada
Catalog number 04422058164; – Nordics
Catalog number 04422040170. (Distributed
outside of the U.S.) Recall # Z-0424-06;
i) Roche brand Accu-Chek Aviva Meter (mg/dL);
Catalog number 03532356001. (Distributed
outside of the U.S.) Recall # Z-0425-06;
j) Roche brand Accu-Chek Aviva Meter (mmol/L);
Catalog numbers 03532313001 and 03532321001.
(Distributed outside of the U.S.)
Recall # Z-0426-06

REASON
Incorrect patient blood glucose test results may be reported due to a software problem.

CODE
a), b), c), d),e), f), g), Aviva meter serial
numbers 52500000000 through 52510999999;
h), i) and j) Aviva meter serial numbers
52600000000 through 52610999999, 52700000000
through 52710999999 and 52800000000 through
52810999999

RECALLING FIRM/MANUFACTURER
Recalling Firm: Roche Diagnostics Corp., Indianapolis, IN, by letters dated November 9, 2005 and press release issued on January 17, 2006.
Manufacturer: Sanmina SCI, Huntsville, AL. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
1,143,155 meters

DISTRIBUTION
Nationwide and Internationally

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.