Inc. Class I Beckman Coulter

Company: Beckman Coulter, Inc., .
Date of Enforcement Report: 9/6/2005
Class:I

PRODUCT
Synchron LX Systems Version 4.5 Operating Software, Recall # Z-0009-06

REASON
Synchron LX Systems Operating Software Version 4.5/v4.6 CBD (Chemistry Database) introduced an adjustment to the urine Uric Acid database calculation scheme which for some patients, depending on the volume of 24 hour urine sample, results may be reduced from above the upper limit of the laboratory’s reference interval to below the limit.

CODE
Synchron LX 20 and PRO Version 4.5 Operating Software Part Number A21127 (Upgrade Kit) Software Versions V4.5N4.6 CDB (Chemistry Database). Synchron LXi Version 4.5 Operating Software A21128 (Upgrade Kit) Software Versions V4.5N4.6 CDB(Chemistry Database)

RECALLING FIRM/MANUFACTURER
Beckman Coulter, Inc., Brea, CA, by letter dated September 6, 2005. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
868 units
DISTRIBUTION
Nationwide and Canada

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.