Medtronic Surgical Navigation Technologie Class II

Company: Medtronic Surgical Navigation Technologies, Inc.
Date of Enforcement Report: 4/20/05
Class: II

PRODUCT
Stealth Station Cranial Application v.4.5. Catalog Numbers: 960-201 and 9730888. Recall # Z-0711-05.

REASON
The Stealth Station Cranial Application Version 4.5 contains a software anomaly, which may present a hazard to the patient. The user has the option to create and make visible supplemental contours of specific anatomy viewable during a surgical procedure. Activation of a new tip extension feature will move the location of these contours in such a manner that they no longer correlate to the intended anatomy. Potential consequences may include improper resection of the brain tissue.

CODE
Software v.4.5.

RECALLING FIRM
Medtronic Surgical Navigation Technologies, Inc., Louisville, CO, by letters dated January 28, 2005 and February 2, 2005. Firm initiated recall is complete.

VOLUME OF PRODUCT IN COMMERCE
44 systems.

DISTRIBUTION
Nationwide and Internationally.

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.