Day

March 22, 2005
Recipient: Tosoh Bioscience Inc.Product: Immunoassay diagnostic packs and immunoassay systems Date: 3/22/05 Specifically, our inspections revealed that on or about March 27, 2003, your firm was notified by Tosoh Corporation, Tokyo, Japan that a new software revision, version 3.12, was being released to correct washing errors in the analyzer that resulted in discrepant troponin results....