Company: Baxter Healthcare Corp.
Date of Enforcement Report: 1/12/05
 Class:  II
PRODUCT
 a) Colleague Single Channel Volumetric
 Infusion Pumps, product code 2M8151
 and 2M8151R; Made in Singapore; monochrome
 display screen. Recall # Z-0342-05;
 b) Colleague 3 Triple Channel Volumetric
 Infusion Pumps, product code 2M8153
 and 2M8153R; Made in Singapore; monochrome
 display screen. Recall # Z-0343-05;
 c) Colleague CX Single Channel Volumetric
 Infusion Pumps, product code 2M8161
 and 2M8161R; Made in Singapore; color
 display screen. Recall # Z-0344-05;
 d) Colleague 3 CX Triple Channel Volumetric
 Infusion Pumps, product code 2M8163 and
 2M8163R; Made in Singapore; color
 display screen. Recall # Z-0345-05.
 REASON
 A software anomaly causes a failure alarm code 12:303 which audibly alarms and stops the function of all channels in use. This causes an interruption in therapy, which may result in a risk to the patient.
 CODE
 All serial numbers.
 RECALLING FIRM/MANUFACTURER
 Baxter Healthcare Corp., Round Lake, IL, by letters dated November 4, 2004. Firm initiated recall is ongoing.
 VOLUME OF PRODUCT IN COMMERCE
 199,456 units.
 DISTRIBUTION
 Nationwide and Internationally.
