Baxter Healthcare Renal Div Class II

Company: Baxter Healthcare Renal Div
Date of Enforcement Report: 5/19/04
Class: II

PRODUCT
a) Baxter Renal Software Suite (Renal Link and PD Link).
Recall # Z-0901-04;
b) Renalsoft software system. Recall # Z-0902-04.

REASON
There are software anomalies present that may increase the possibility of a medication error if the dialysis clinic software is used to track, record, or coordinate the administration of medications.

VOLUME OF PRODUCT IN COMMERCE
4,164 units.
CODE
a) Product codes 5M5590, 5M5591, 5M5592, 5M5593, 5M5594,
5M5595, 5M5596, 5M5603, 5M5604, R5C4532, R5C4534,
5M5597, 5M5541, 5M5541R, all software versions;
b) Product code RENLSFT, all software versions.

RECALLING FIRM/MANUFACTURER
Baxter Healthcare Renal Div., McGaw Park, IL, by letters dated April 12, 2004. Firm initiated recall is ongoing.

DISTRIBUTION
Nationwide.

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.