Cordis Corporation

Recipient:Cordis Corporation
Product: Sirolimus-Eluting Coronary Stent
Date: 4/1/2004

Failure to validate with a high degree of assurance, processes, including changed processes, that cannot be fully verified by subsequent investigation and test, as required by 21 CFR 820.75(a) & (c). For example:
The automated [redacted] data acquisition system, used to ensure the integrity of the analytical data generated from laboratory chromatography equipment, was not adequately validated for its intended use. The validation did not include testing and verification of backup and restoration of the electronic data files.

Issuing Office: CDRH

Upcoming Training

62304, FDA, and Emerging Standards for Medical Device and HealthIT
Instructors:  Brian Pate, John F. Murray, Jr
Location: Sunnyvale, CA, USA
Dates:  February 4-6, 2020
Registration Link

Receive $300 discount with Premium-Individual subscription purchase (or $333 per person for Premium-Company subscription)!  Email training@softwarecpr.com
to receive discount