CSI Spectroscopy Class II

REASON

Due to a software anomaly, the device acquires data from a slice thicker than intended, and may not be at isocenter.

PRODUCT
Chemical-Shift Imaging (CSI) Spectroscopy, Model 90LNI.

Recall #Z-412-0.

CODE

Serial Numbers: AF212E, BA109P, BA218P, BA103P, AF165E, AF157EU, BA183P,
BA118P, BA182P, BA164D, BA178P, BA114P, BA225P, BA187P, AF113EU, BB104P,
BA110P, BA131D, and BA192D.

MANUFACTURER

Marconi Medical Systems, Inc., Highland Heights, Ohio.

RECALLED BY

Manufacturer, by letter dated February 3, 2000. Firm-initiated recall ongoing.

DISTRIBUTION

Florida, Maryland, Massachusetts, Utah, Arizona, Australia, Italy, Germany, Japan, China and England.

QUANTITY

19 units were distributed.